EasyManua.ls Logo

Tektronix TDS 620A - Page 273

Tektronix TDS 620A
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–255
TRIGger:MAIn:PULse:RUNT:THReshold:HIGH?
TRIGger
:
:
PULse
MAIn
RUNT
:
:
?
<Space>
<NR3>
HIGH
THReshold
<NR3> is the threshold, in volts.
TRIGGER:MAIN:PULSE:RUNT:THRESHOLD:HIGH 120E-3
sets the upper limit of the pulse runt trigger to 120 mV.
TRIGger:MAIn:PULse:RUNT:THReshold:LOW
TDS 5XXA, 6XXA, & 7XXA Only
Sets or queries the lower limit for the main pulse runt trigger. This command is
equivalent to setting the threshold in the Pulse Runt Threshold side menu.
Trigger
TRIGger:MAIn:PULse:RUNT:THReshold:LOW <NR3>
TRIGger:MAIn:PULse:RUNT:THReshold:LOW?
THReshold
TRIGger :
:
PULse
MAIn
RUNT :
:
?
<Space>
<NR3>
LOW
<NR3> is the threshold, in volts.
TRIGGER:MAIN:PULSE:RUNT:THRESHOLD:LOW 50E-3
sets the lower limit of the pulse runt trigger to 50 mV.
Arguments
Examples
Group
Syntax
Arguments
Examples

Table of Contents

Other manuals for Tektronix TDS 620A

Related product manuals