Command Descriptions
2–256
TDS Family Oscilloscope Programmer Manual
TRIGger:MAIn:PULse:RUNT:WHEn
TDS 684A & 7XXA Only
Sets or queries the type of pulse width the trigger checks for when it uncovers a
runt. This is equivalent to to selecting Trigger When from the main Trigger’s
Pulse, Runt menu and Occurs or Wider Than in the resulting side menu.
Trigger
# " $ ! %
"
!
" specifies a trigger if a runt of any detectable width occurs.
! specifies a trigger if a runt of greater than the specified width occurs.
sets the runt trigger to occur when the oscilloscope detects a runt in a pulse wider
than the specified width.
TRIGger:MAIn:PULse:RUNT:WIDth
TDS 684A & 7XXA Only
Sets or queries the minimum width for a valid main pulse runt trigger. This
command is equivalent to entering a value in the Trigger menu’s Wider Than
side menu.
Trigger
!
!
Group
Syntax
Arguments
Examples
Group
Syntax