Saving Waveforms and Setups
TDS 500B, TDS 600B, & TDS 700A User Manual
3–117
2. To recall a setup stored on disk, press From File (side). Then use the general
purpose knob to select the exact file from the resulting scrollbar list. Only
files with .set extensions will be displayed. Finally, press Recall From
Selected File (side) to complete the operation.
Recalling a setup will not change the menu that is currently displayed. If you recall
a setup that is labeled factory in the side menu, you will recall the factory setup.
(The conventional method for recalling the factory setup is described below.)
To reset your oscilloscope to the factory defaults:
Press SAVE/RECALL SETUP Recall Factory Setup (main) OK
Confirm Factory Init (side).
See Factory Initialization Settings, on page D–1, for a list of the factory defaults.
Sometimes you might use the oscilloscope to acquire waveforms that are
confidential. Furthermore, before returning the oscilloscope to general usage,
you might want to remove all such waveforms and any setups used to acquire
them. (Be sure you want to remove all waveforms and setups, because once they
are removed, you cannot retrieve them.) To use Tek Secure to remove all
reference setups and waveforms (does not affect mass storage disk):
Press SHIFT UTILITY System (main) Config (pop-up) Tek Secure
Erase Memory (main) OK Erase Setup & Ref Memory (side).
Executing Tek Secure accomplishes the following tasks:
H Replaces all waveforms in reference memories with zero sample values.
H Replaces the current front panel setup and all setups stored in setup memory
with the factory setup.
H Calculates the checksums of all waveform memory and setup memory
locations to verify successful completion of setup and waveform erasure.
H If the checksum calculation is unsuccessful, displays a warning message; if
the checksum calculation is successful, displays a confirmation message.
To run file utilities, see the Managing the File System on page 3–121.
See Example 4: Saving Setups, on page 2–24.
See Appendix D: Factory Initialization Settings on page D–1.
To Recall the
Factory Setup
To Delete All Setups and
Waveforms — Tek
Securer
To Run the File Utilities
To Find More Information

Table of Contents

Other manuals for Tektronix TDS 684A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 684A and is the answer not in the manual?

Tektronix TDS 684A Specifications

General IconGeneral
BrandTektronix
ModelTDS 684A
CategoryTest Equipment
LanguageEnglish

Related product manuals