Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−57
Refer to Registers on page 3-1 for more information.
Status and Error
*CLS
DESE, *ESE, *ESR?, EVENT?, EVMsg?, *SRE, *STB?
CURSor? (Query Only)
Returns all current oscilloscope cursor settings.
Cursor
CURSor?
Oscilloscope cursor parameters
CURSor?
might return CURSOR:FUNCTION VBARS;SELECT:SOURCE CH1;
:CURSOR:VBARS:UNITS SECONDS;POSITION1 0.0E0;POSITION2
4.48E–3;:CURSOR:HBARS:UNITS VOLTS;POSITION1 0.0E0;POSI−
TION2 5.0E–1 as the current cursor settings.
CURSor:FUNCtion
Selects and displays the oscilloscope cursor type. Cursors are
attached to the waveform selected by CURSor:SELect:SOUrce. This
command is equivalent to setting the Type option in the Cursor
Group
Syntax
Related Commands
Group
Syntax
Returns
Examples

Table of Contents