Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−99
HORizontal:MAIn:SCAle <NR3>
HORizontal:MAIn:SCAle?
<NR3> is the time per division. The range depends on the oscillo-
scope model. The acceptable values are in a 1–2.5–5 sequence.
Other values are forced to the closest acceptable value.
HORizontal:MAIn:SCAle 2.5E–6
sets the main scale to 2.5 ms per division.
HORizontal:MAIn:SECdiv
Sets the time per division for the oscilloscope main time base. This
command is identical to the HORizontal:MAIn:SCAle command. It
is provided to maintain program compatibility with some older
models of Tektronix oscilloscopes.
HORizontal:POSition
Sets or queries the main time base horizontal position. This
command is identical to the HORizontal:MAIn:POSition command.
This command is included for compatibility.
Horizontal
HORizontal:POSition <NR3>
HORizontal:POSition?
<NR3> is the position in seconds. This value is the difference between
the trigger point and the center graticule. Positive values place the
trigger before the center graticule.
Syntax
Arguments
Examples
Group
Syntax
Arguments