Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−1 13
MEASUrement:IMMed:SOUrce
Sets or queries the source for all immediate measurements.
Measurement
MEASUrement:IMMed:SOUrce CH<x>
MEASUrement:IMMed:SOUrce?
CH<x> is an input channel.
MEASUrement:IMMed:SOUrce CH1
specifies channel 1 as the immediate measurement source.
MEASUrement:IMMed:TYPe
Sets or queries the immediate measurement type.
Measurement
MEASUrement:IMMed:TYPe { FREQuency | MEAN | PERIod |
PK2pk | CRMs | MINImum | MAXImum | RISe | FALL | PWIdth
| NWIdth }
MEASUrement:IMMed:TYPe?
FREQuency is the reciprocal of the period measured in Hertz.
MEAN is the arithmetic mean over the entire waveform.
Group
Syntax
Arguments
Examples
Group
Syntax
Arguments