Command Descriptions
2−122
TDS200/1000/2000 Series Oscilloscope Programmer Manual
Status and Error
*OPC
*OPC?
Using the *OPC? query to check for completion of a single sequence
acquisition:
ACQuire:STOPAfter SEQUENCE
ACQuire:STATE ON
*OPC? will return a 1 after the acquisition is complete.
Refer to page 3-14 for examples of how to use the *OPC set
command.
ALLEv, BUSY?, DESE, *ESR?, *WAI
*PSC
Sets and queries the power-on status flag that controls the automatic
power-on handling of the DESER, SRER, and ESER registers. When
*PSC is true, the DESER register is set to 255 and the SRER and
ESER registers are set to 0 at power on. When *PSC is false, the
current values in the DESER, SRER, and ESER registers are
preserved in nonvolatile memory when power is shut off and are
restored at power on. Refer to the Status and Events chapter on
page 3-1 for more information.
Status and Error
*PSC <NR1>
*PSC?
Group
Syntax
Examples
Related Commands
Group
Syntax