Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−123
<NR1> = 0 sets the power-on status clear flag to false, disables the
power on clear, and allows the oscilloscope to possibly assert SRQ
after power on.
<NR1> 0 sets the power-on status clear flag true. Sending *PSC 1,
therefore, enables the power-on status clear and prevents any SRQ
assertion after power on.
*PSC 0
sets the power-on status clear flag to false.
*PSC?
might return the value 1, showing that the power-on status clear flag
is set to true.
DESE, *ESE, FACtory, *RST, *SRE
*RCL (No Query Form)
Restores the state of the oscilloscope from a copy of its settings
stored in memory. (The settings are stored using the *SAV
command.) This command is equivalent to RECAll:SETUp, and
performs the same function as the Recall Saved Setup item in the
front-panel Save/Recall Setup menu.
Save and Recall
*RCL <NR1>
<NR1> is an integer value in the range from 1 to 5 (TDS200 series) or
1 to 10 (TDS1000 and TDS2000 series), and specifies a setup storage
location.
Arguments
Examples
Related Commands
Group
Syntax
Arguments