Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−83
<NR1> is a value in the range from 0 through 255. The binary bits of
the ESER are set according to this value.
The power-on default for ESER is 0 if *PSC is 1. If *PSC is 0, the
ESER maintains its value through a power cycle.
NOTE . Setting the DESER and the ESER to the same value allows
only those codes to be entered into the Event Queue and summarized
on the ESB bit (bit 5) of the Status Byte Register. Use the DESE
command to set the DESER. Refer to Event Handling Sequence on
page 3-8 for more information.
*ESE 209
sets the ESER to binary 11010001, which enables the PON, URQ,
EXE, and OPC bits.
*ESE?
might return the string *ESE 186, showing that the ESER contains
the binary value 10111010.
*CLS, DESE, *ESR?, EVENT?, EVMsg? *SRE, *STB?