Command Descriptions
2−84
TDS200/1000/2000 Series Oscilloscope Programmer Manual
*ESR? (Query Only)
Returns the contents of the Standard Event Status Register (SESR).
*ESR? also clears the SESR (since reading the SESR clears it).
Refer to the Status and Events chapter on page 3-1 for more
information.
Status and Error
*ESR?
Contents of the Standard Event Status Register.
*ESR?
might return the value 213, showing that the SESR contains binary
11010101.
ALLEv?, *CLS, DESE, *ESE, EVENT?, EVMsg?, *OPC, *SRE,
*STB?
EVENT? (Query Only)
Returns from the Event Queue an event code that provides
information about the results of the last *ESR? read. EVENT? also
removes the returned value from the Event Queue. Refer to Event
Handling Sequence on page 3-8 for for more information.
Status and Error
EVENT?
Group
Syntax
Returns
Examples
Related Commands