EasyManua.ls Logo

Tektronix TDS2012 - Page 92

Tektronix TDS2012
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Descriptions
2−74
TDS200/1000/2000 Series Oscilloscope Programmer Manual
<NR1> = 2 sets the number of bytes per waveform data point to
2 bytes (16 bits). If DATa:WIDth is set to 2, the least significant byte
is always zero.
DATa:WIDth 1
sets the data width to 1 byte per data point for CURVe data.
CURVe, WFMPre:BIT_Nr, WFMPre:BYT_Nr
*DDT
Lets you specify a command or a list of commands to execute when
the oscilloscope receives a *TRG command or the GET GPIB
interface message. This is a special alias that *TRG uses.
Miscellaneous
*DDT { <Block> | <QString> }
*DDT?
<Block> or <QString> is a complete sequence of program messages.
The messages must contain only valid commands that must be
separated by semicolons and must follow all rules for concatenating
commands (refer to page 2-6). The sequence must be 80 characters.
<Block> format is always returned as a query response.
*DDT #217ACQuire:STATE RUN<EOI>
specifies that the acquisition system will be started each time a
*TRG command is sent.
Examples
Related Commands
Group
Syntax
Arguments
Examples

Table of Contents

Other manuals for Tektronix TDS2012