Command Descriptions
2−76
TDS200/1000/2000 Series Oscilloscope Programmer Manual
DESE 209
sets the DESER to binary 11010001, which enables the PON, URQ,
EXE, and OPC bits.
DESE?
might return the string :DESE 186, showing that DESER
contains the binary value 10111010.
*CLS, *ESE, *ESR?, EVENT?, EVMsg?, *SRE, *STB?
DIAg:RESUlt:FLAg? (Query Only)
Returns the Pass/Fail status from the last diagnostic test sequence
execution (those run automatically at power on, or those requested
through the Service Menu). Use the DIAg:RESUlt:LOG? query to
determine which test(s) has failed.
Calibration and Diagnostic
DIAg:RESUlt:FLAg?
PASS means that the oscilloscope passes all diagnostic tests.
FAIL means that the oscilloscope has failed at least one of the
diagnostic tests.
DIAg:RESUlt:FLAG?
returns either PASS or FAIL.
Examples
Related Commands
Group
Syntax
Returns
Examples