EasyManuals Logo
Home>Tektronix>Test Equipment>TDS2022

Tektronix TDS2022 User Manual

Tektronix TDS2022
207 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #145 background imageLoading...
Page #145 background image
Math FFT
126
TDS1000/2000-Series Digital Oscilloscope User Manual
Measuring an FFT Spectrum Using Cursors
You can take two measurements on FFT spectrums: magnitude
(in dB) and frequency (i n Hz). Magnitude is referenced to 0 dB,
where 0 dB equals 1 V
RMS
. You can use the cursors to take
measurements at any zoom factor.
Push CURSOR
" Source and select Mat h. Push the Type option
button to select between Magnitude and Frequency. Use the Vertical
Position knobs to move cursors 1 and 2.
Use hori zontal cursors to me asure magnitude and vertical cursors to
measure frequency. The options display the delta between the two
cursors, the value at cursor 1 position, and the value at cursor 2
position. Delta is the absolute value of cursor 1 minus cursor 2.
Magnitude cursors
Frequency cursors
You can also take a frequency measurement. To do so, turn the
Horizontal Position knob to position a frequency component on the
center graticule line and rea d the frequency at the top right of the
display.

Table of Contents

Other manuals for Tektronix TDS2022

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2022 and is the answer not in the manual?

Tektronix TDS2022 Specifications

General IconGeneral
BrandTektronix
ModelTDS2022
CategoryTest Equipment
LanguageEnglish

Related product manuals