Performance Verication
Test Record
Table 13: Test record
Instrument Serial Number:
Temperature
:
Date of Calibration:
Certicate Number:
RH %:
Technician:
Instrument
performance test Minimum Incoming Outgoing Maximum
5 mV/div
33.6 mV 36.4 mV
200 mV/div
1.358 V 1.442 V
Channel 1
DC Gain Ac curac y
2V/div
13.58 V 14.42 V
5 mV/div
33.6 mV 36.4 mV
200 mV/div
1.358 V 1.442 V
Channel 2
DC Gain Ac curac y
2V/div
13.58 V 14.42 V
5 mV/div
33.6 mV 36.4 mV
200 mV/div
1.358 V 1.442 V
Channel 3
DC Gain Accuracy
1
2V/div
13.58 V 14.42 V
5 mV/div
33.6 m
V
36.4 m
V
200 mV/div
1.358
V
1.442
V
Channel 4
DC Gain Accuracy
1
2V/div
13.5
8V
14.4
2V
Channel 1 Bandwidth
2.12
V
2
Channel 2 Bandwidth
2.1
2V
2
Channel 3 Bandwidth
1
2.1
2V
2
Channel 4 Bandwidth
1
2.
12 V
2
Sample Rate and Delay Time Accuracy
-
2divs
+
2divs
Channel 1 Edge Trigger Sensitivity Stable trigger
3
Channel 2 Edge Trigger Sensitivity Stable trigger
3
Channel 3 Edge Trigger Sensitivity
1
Stable trigger
3
Channel 4 Edge Trigger Sensitivity
1
Stable trigger
3
External Edge Trigger Sensitivity Stable trigger
3
Channel 1 Vertical Position Accuracy,
Minimum margin
0
Channel 2 Vertical Position Accuracy,
Minimum margin
0
Channel 3 Vertical Position Accuracy
1
,
Minimum margin
0
Channel 4 Vertical Position Accuracy
1
,
Minimum margin
0
1
Channels 3 and 4 are only available on four channel oscilloscopes.
2
The bandwid th te st do es not have a high limit.
3
Thelimitsvarybymodel. Checkthe procedure for the correct limits.
28 TDS2000C and TDS1000C-EDU Series Oscilloscope Service Manual