Incoming Inspection
Simultaneous Input
Operation Test (Option
SIM)
1. Set up a signal g
enerator with an HDVG1 module and a DVG1 module.
Connect an output from the HDVG1 module to the SDI A input on the
instrument under test.
2. Connect an output from the DVG1 module to the SDI B input on the
instrument under test.
3. Connect the loopthrough terminators on the instrument.
4. Set the HDVG
1 module format to 1080 59.94i 100% color bars.
5. Set the DVG1 module format to 525-270.
6. Press FACTORY to restore factory settings.
7. Press CONFIG, select Input Mode, and then select Simultaneous.
8. Press CONFIG again to remove the menu.
9. Press DISPLAY SELECT button 2 to select tile 2.
10. Press t
he WFM button to select the waveform display in tile 2.
11. Press DISPLAY SELECT button 4 to select tile 4.
12. Press the PICT button to select the picture display.
13. Check that the left two tiles (tiles 1 and 3) show the signal from the SDI
A input. Check that the right two tiles (tiles 2 and 4) show the signal from
the SDI B input.
14. Check that the left portion of the Status Bar displays the status of the SDI A
signal input, and that the right portion of the Status Bar displays the status of
the
SDI B signal input.
15. Enter Pass or fa il i n the test record.
Waveform Rasterizers Specications and Performance Verication 6 9

Table of Contents

Other manuals for Tektronix WVR6020

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix WVR6020 and is the answer not in the manual?

Tektronix WVR6020 Specifications

General IconGeneral
BrandTektronix
ModelWVR6020
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals