Incoming Inspection
Dual Link Format
Verication Test (Option
DL)
1. Set up a dual lin
k signal generator with an HDLG7 module. Connect the Link
A output on the module to the SDI A input on the instrument, and the Link B
output on the module to the SDI B input on the instrument.
2. Connect the loopthrough terminators on the instrument.
3. Perform the following steps to set the HDLG7 module format to 1080 59.94i:
a. On the signa
l generator, press the FORMAT button until 1080 59.94i
appears, and then press the ENTER button.
b. Press the C
ANCEL button.
c. Press the up/down arrows until Sample Structure/Depth appears.
d. Press the left/right arrows until YCbCr+A 4:4:4 10 bits appears.
e. Press the ENTER button.
f. Press the COLOR BAR button until 75% color bars appears.
4. Press the FACTORY button on the instrument under test to restore factory
settings.
5. Press DISPLAY SELECT 2 to select tile 2.
6. Press
andholdtheSTATUS button.
7. Select Display Type > Video Session.
8. Press DISPLAY SELECT 4 to select tile 4.
9. Press and hold the VECT button to display the Vector map and the Vector
menu.
10. Select Display Type > Lightning.
11. Press the VECTOR button to close the menu.
70 Waveform Rasterizers Specications and Performance Verication

Table of Contents

Other manuals for Tektronix WVR6020

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix WVR6020 and is the answer not in the manual?

Tektronix WVR6020 Specifications

General IconGeneral
BrandTektronix
ModelWVR6020
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals