EasyManua.ls Logo

TEL AFA4000/1 - 12 Operation: Auto Sash Controller; Control Philosophy; Modes of Operation

TEL AFA4000/1
175 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AFA4000/1 Airflow Monitor / p.143
12. Operation: Auto Sash Controller
12.1 Control philosophy
When all components are installed, setup and calibrated correctly, the Auto Sash
Controller closes the sash after a pre-set delay, provided that the area in front of the
fume hood is clear.
The following additional protocols also apply:
When the user returns to the fume hood during the delay period, the control timer
will stop and reset once the user leaves the fume hood again.
When the user returns to a fume hood that is closing, the drive will stop when it
senses the user.
When an obstruction is detected under the sash, the Auto Sash Controller will not
open or close the sash and will give a visual alarm to indicate a fault condition.
When an obstruction is detected whilst the sash is closing, the sash drive will stop and
give a visual indication to demonstrate a fault condition. To reset the fault alarm,
press the reset button.
When the sash starts to close a timer is initiated. When the sash does not close within
the specified time, the sash drive will stop and will give a visual alarm to indicate a
fault condition.
When a user is present in front of the fume hood, the sash can be operated by hand.
12.2 Modes of operation
The Auto Sash Controller has the following modes of operation:
User Present
User Not Detected
Auto-Open (User Configurable)
Building Management System (BMS) Inputs
Sash Lock

Table of Contents