Overview
This instruction sheet describes the steps to install
the optional pump heater inside the pump cover of both
the 5800 and 4700 Samplers. The heater prevents liquid
from freezing inside the pump under extremely cold con-
ditions.
The heater prevents liquid from freezing only inside
the pump housing cover. In extremely cold climates,
you must also route the suction line correctly to
prevent liquid from freezing in the suction line. The
suction line must have a continuous slope. Do not
allow loops or low spots in the line where liquid can
remain between samples. (This general practice
applies to all climates to prevent sample
cross-contamination.)
Required Parts and Tools
● Heater Installation Kits
❍ 68-4700-104 for 115 VAC Samplers
❍ 68-5800-060 for 230 VAC Samplers
● #2 Phillips screwdriver
● Hack saw
● Fine sand paper or file
Installation
The Heater Installation Kit includes a heater, a heat
shield, self-tapping screws, and a clip to secure the AC
line cord. To install the kit:
1. Refer to Figure 1. If a notch is not present, cut a
notch in the pump cover with a hack saw or sim-
ilar tool. Use sand paper or a file to remove burrs
and smooth the cut edge.
Figure 1: Cut a notch in the top cover
for the AC line
cord to pass through
2. Fit the heat shield inside the pump cover as
shown in Figure 2. Secure the shield with the
four #6
3
/8" self-tapping screws.
Figure 2: Install heat shield inside cover
Instruction Sheet 69-4703-150
Revision D, February 1, 2016