EasyManua.ls Logo

Tenma 72-7630 - Channel Bandwidth Setup; Vertical Volts;Div Adjustment; Probe Rate Setup; Waveform Inversion Setup

Tenma 72-7630
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
Channel bandwidth setup
If for example a signal is applied to CH1
which is a pulse signal that contains high
frequency oscillation.
Press CH1 to select Channel1.
Press F2 to set the BANDWIDTH LIMIT
OFF so it is set up as full bandwidth.
The signal being measured can now pass
through even if it contains high frequency quantities.
Press F2 to set BANDWIDTH LIMIT ON
so that frequency quantities higher than
20MHz in the signal being measured will
be limited.
Probe rate setup
To match the probe attenuation factor
setup, it is necessary to set up the probe
attenuation factor in the channel operation
menu accordingly.
For example when the probe attenuation
factor is 10:1, set the probe attenuation
factor at 10X in the menu. This principle
applies to other values to ensure the voltage reading is correct.
Waveform inversion setup
The displayed signal is inverted 180 degrees
with respect to the
ground level.
Vertical Volts/Div adjustment setup
The VOLTS/DIV range of the vertical
deection factor can be adjusted either in
coarse or ne tune mode.
In COARSE TUNE the VOLTS/DIV range is
2mV/div~5V/div. Tuning is in steps of 1-2-5.
In FINE TUNE mode the deection factor
can be adjusted in smaller steps allowing
continuous adjustment within the range
2mV/div~5V/div without interruption.

Related product manuals