Storage temperature :
and humidity –10 to 60 (14 to 140),
below 70%RH.
Dimensions : 170(L)×52(W)×38mm(H)
6.7(L)×2.1(W)×1.5(H) inches.
Weight : Approx. 180g with battery.
Accessories : instruction manual, carry case,
battery.
3-2 Electrical Specifications
Infrared
Measuring range : -20 to 500 (-4 to 932)
Resolution : 1 , 1
Accuracy : ± 3% reading or ± 3 or 6.
Temperature
coefficient : 0.1 times the applicable accuracy
specification per from 0 to 18
and 28to 40(32to 64and 82
to 104).
Responding time : once per second.
Spectral Response : 6 ~ 14um.
Field of view : 8:1 ; optics ratio with a 1” min target.
Emissivity : 0.1 ~ 1.0
Sighting : Laser marker 1mw (class 2).
Sensor : Thermopile.
4
Type K
Measuring range : -50 to 1333 (-58 to 1999).
Resolution : 0.1 , 1 , 0.1 , 1.
Responding time : once per second.
Basic accuracy : (@ 23±5 calibration) accuracy are ±
(…% of reading + degree) at 18 to
28 with relative humidity up 80%.
Function Resolution Range Accuracy
-50
℃〜
0
±
( 0.2%rdg + 1.0
)
0.1
0
℃〜
200
±
( 0.1%rdg + 0.8
)
1
200
℃〜
1333
±
( 0.2%rdg + 2
)
-58
℉〜
32
±
( 0.2%rdg + 2
)
0.1
32
℉〜
200
±
( 0.1%rdg + 1.6
)
1
200
℉〜
1999
±
( 0.2%rdg + 3
)
Temperature Coefficient:
0.1 times the applicable accuracy specification per
from o to 18 and 28 to 40 ( 32 to 64 and
82 to 104 ).
Note :
The basic accuracy specification does not include the
error of the probe. Please refer to the probe
accuracy specification for additional details.