Storage temperature :
and humidity –10℃ to 60℃ (14℉ to 140℉),
below 70%RH.
Dimensions : 170(L)×52(W)×38mm(H)
6.7(L)×2.1(W)×1.5(H) inches.
Weight : Approx. 180g with battery.
Accessories : instruction manual, carry case,
battery.
3-2 Electrical Specifications
Infrared
Measuring range : -20℃ to 500℃ (-4℉ to 932℉)
Resolution : 1℃ , 1℉
Accuracy : ± 3% reading or ± 3℃ or 6℉.
Temperature
coefficient : 0.1 times the applicable accuracy
specification per ℃ from 0℃ to 18℃
and 28℃to 40℃(32℉to 64℉and 82℉
to 104℉).
Responding time : once per second.
Spectral Response : 6 ~ 14um.
Field of view : 8:1 ; optics ratio with a 1” min target.
Emissivity : 0.1 ~ 1.0
Sighting : Laser marker 1mw (class 2).
Sensor : Thermopile.
4
Type K
Measuring range : -50℃ to 1333℃ (-58℉ to 1999℉).
Resolution : 0.1℃ , 1℃ , 0.1℉ , 1℉.
Responding time : once per second.
Basic accuracy : (@ 23±5℃ calibration) accuracy are ±
(…% of reading + degree) at 18℃ to
28℃ with relative humidity up 80%.
Function Resolution Range Accuracy
-50
℃〜
0
℃
±
( 0.2%rdg + 1.0
℃
)
0.1
℃
0
℃〜
200
℃
±
( 0.1%rdg + 0.8
℃
)
℃
1
℃
200
℃〜
1333
℃
±
( 0.2%rdg + 2
℃
)
-58
℉〜
32
℉
±
( 0.2%rdg + 2
℉
)
0.1
℉
32
℉〜
200
℉
±
( 0.1%rdg + 1.6
℉
)
℉
1
℉
200
℉〜
1999
℉
±
( 0.2%rdg + 3
℉
)
Temperature Coefficient:
0.1 times the applicable accuracy specification per ℃
from o℃ to 18℃ and 28℃ to 40℃ ( 32℉ to 64℉ and
82℉ to 104℉ ).
Note :
The basic accuracy specification does not include the
error of the probe. Please refer to the probe
accuracy specification for additional details.
5