8 Technical data
100
8.2.4 testo Saveris T3 / T3D
2-channel radio data logger with two external TC probe connections (TC
characteristics can be selected).
Feature Value
Probe type
TC
Measuring range
TC type J
TC type K
TC type S
TC type T
-100 to +750°C
-195 to +1350°C
0 to +1760°C
-200 to +400°C
Accuracy
±0.5°C or 0.5% of measured value (25°C)
Resolution
0.1°C / TC type S 1°C
Connection
2 TCs via TC socket,
max. difference in potential 2 V
Protection class
IP 54 (with probe inserted or connection socket
sealed with rubber plug)
Operating temperature
-20 to +50°C
The probe inputs are not galvanically isolated from one another. Take
this into account when using probes with a non-isolated thermocouple.
8.2.5 testo Saveris Pt / PtD
Radio data logger with an external Pt100 probe connection.
Feature Value
Probe type
Pt100
Measuring range
-200 to +600°C