Low Temp Incubator 3-3Thermo Scientific
Section 3
Preventive Maintenance
Preventive Maintenance for Model 815/818 Incubators
Refer to
Manual
Section
Action Daily Weekly Yearly
3 Check temperature display versus setpoints, w/ independent instrument
9
-- Inspect door latch, hinges and door gasket seal.
9
3 Verify and document temp calibration, as applicable. See Calibration section.
9
5 Perform a complete decontamination procedure. Wipe down interior, shelves
and side panels with disinfectant. Rinse everything well with sterile distilled
water.
Between experiments
More frequent decontamination
may be required, depending on use
and environmental conditions
PREVENTIVE MAINTENANCE
Incubators
Your equipment has been thoroughly tested and calibrated before shipment. Regular preventive maintenance is important to keep your unit
functioning properly. The operator should perform routine cleaning and maintenance on a regular basis. For maximum performance and
efficiency, it is recommended the unit be checked and calibrated periodically by a qualified service technician.
The following is a condensed list of preventive maintenance requirements. See the specified section of the operating manual for further
details.
We have qualified service technicians, using NIST traceable instruments, available in many areas. For more information on Preventive
Maintenance or Extended Warranties, please contact us at the number listed below.
Cleaning and calibration adjustment intervals are dependent upon use, environmental conditions and accuracy required.
Tips for all incubators:
x Do NOT use bleach or any disinfectant that has high chloros
x Use sterile
, distilled or demineralized water.
x Do not use powdered gloves for tissue cultures.