The following table lists the available commands for monitoring or
controlling your unit.
To View Data: If Status Update Time is “0.0” no data will automatically be
displayed. To view data within a particular command code, type that
command code followed by the question mark (?) symbol and then
ENTER (Return).
Example: To view Setpoint Temperature type: T? <CR>
To Change Data: To change the value within the command type the
command code followed by the new value and ENTER (Return).
Example 1: To change Setpoint Temperature to 25.0°C type: T25.0 <CR>
Example 2: To change Setpoint Temperature to -.8°C type: T-.8<CR>
(A) Actual Temperature: The actual temperature within the incubator.
(T) Setpoint Temperature: The temperature at which the user has selected
the incubator to operate.
6-6 Low Temp Incubator Thermo Scientific
Section 6
Factory Options
Using the Communication
(continued)
View
Command:
Description: Unit: Range:
Change
Command
A? Actual Temperature Degrees -10 to 50 Read Only
T? Setpoint Temperature Degrees -10 to 50 T10 thru T50
R? Baud Rate - 0 - 5 R0 thru R5
S? Status Update Time - 0 - 8 S0 thru S8
V?
EPROM Version
Number
- Current Rev. Read Only
D? Status Display Type - 0 - 2 D0 thru D2
O?
Display the Alarm
Status
- 0 - 2 Read Onl