What to do if Thermo Scientific ADVANCED has inadequate temperature control?
- MMia RossAug 4, 2025
If you're experiencing inadequate temperature control with your Thermo Scientific Water System, consider the following: * Verify the setpoint. * For refrigerated baths, make sure the condenser is free of dust and debris. * Check the fluid concentration. * Ensure the circulator installation complies with the site requirements. * If the temperature continues to rise, ensure your application's heat load does not exceed the rated specifications. * Enter the controller menu and ensure the ENERGY SAVER mode is on in order for the system to maintain a stable temperature. * Check for high thermal gradients (e.g., the application load is being turned on and off or rapidly changing).