Note Quartz cells MUST be used for all Direct UV 260, Direct UV 280 and
Warburg Christian methods. Glass or plastic cells are not suitable because
they do not transmit fully at the wavelengths used. î
12. Save the results.
Press the Save Results function key on the Results page.
Measure cell growth:
The cell growth measurement uses absorbance at 600nm to indicate the
progress of cell growth in a sample. The instrument does not perform any
calculations or graph the data.
1. Highlight Biological Tests on the Main Menu page.
The Biological Tests sub menu will open.
2. Choose Cell Growth and press Enter.
62 Evolution 300 and Evolution 600 User Guide Thermo Fisher Scientific