c. Setting the REPLACE HEPA filter reminder
A HEPA filter replacement timer can be set for a specific
amount of time, from 1 to 12 months of actual unit running
time. Time will not accrue when the unit is turned off. The
default time is 6 months. When the allotted time has run out,
REPLACE HEPA appears in the display and the visual alarm
flashes. To set the reminder, use the following procedure.
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until REPLACE HEPA XX is dis-
played.
3. Press the up/down arrow to choose the number of
months desired.
4. Press Enter to save the number.
5. Press the Mode key to return to Run Mode or right/left
to go to next/previous parameter.
Note: After the reminder has been set, check the allotted time
remaining by going to Configuration mode, then pressing the
right arrow until NEW HEPA XXX displays. This number is
the remaining days before the filter replacement time specified
runs out. For example, if 12 months was chosen in the
REPLACE HEPA XX message screen, the NEW HEPA number
would be 365 days.
d. Setting an Access Code
A 3-digit Access Code can be entered to avoid unautho-
rized personnel from changing the setpoints, calibration, or con-
figuration. A setting of 000 disables the access code function.
The factory setting is 000.
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until ACCESS CODE XXX is dis-
played in the message center.
3. Press up/down arrow to change the access code.
4. Press Enter to save the access code.
5. Press the Mode key to return to the Run mode or
right/left to go to next/previous parameter.
Section 4 - Configuration
4.1 Configuration Mode
Several features available in the Configuration Mode allow
custom setup of the incubator. These features are listed and
described below. All features may not be necessary in all appli-
cations, but are available if needed. To enter Configuration
mode, press the Mode key until the Configuration indicator
lights. Press the right and/or left arrow until the appropriate
parameter appears in the message center. See Chart 4-1 for
more detail.
a. Turning the Audible Alarm ON/OFF
The audible alarm can be turned on or off. The factory set-
ting is ON.
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until AUDIBLE ON/OFF is dis-
played in the message center.
3. Press up/down arrow to toggle ON or OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left
to go to next/previous parameter.
b. New HEPA Filter
When the REPLACE HEPA reminder is displayed and the
visual alarm flashes, the specified time has elapsed and the
HEPA filter should be replaced.
After replacing the HEPA filter with a new one, clear the
display and reset the timer by following the steps below.
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until NEW HEPA XXX is dis-
played in the message center.
3. Press Enter to restart the timer and clear the REPLACE
HEPA alarm.
4. Press the Mode key to return to Run mode.