Section 10 - Specifications*
*Specifications are based on nominal voltages of 115V or 230V in ambients of 22°C to 25°C.
Temperature
Control ±0.1°C @ 37.0°C
Setpoint Digital touch pad, 0.1°C resolution
Range 5°C over ambient to 50°C, factory default 10.0°C (disabled)
Uniformity ±0.2°C @ +37°C, in a 23°C ambient @ nominal line voltage
Display Digital, LED, 0.1°C increments
Sensor Thermistor
Temperature Safety
Setpoint Fixed @ 160°C, ±5.6°C
Sensor Thermostat, manual reset
Action Removes voltage from heaters to disable heat control
Relative Humidity
Control ±2.0%
Range Ambient level to 95%, non-condensing
Controller Microprocessor
Humidity system 95% RH at 37°C
Setpoint Digital touch pad, 1% resolution
Display Digital LED, 1% resolution
CO
2
Control ±0.1% @ 5.0%
Sensor Infrared
Setpoint Digital touch pad, 0.1°C resolution
Range 0 to 20%, factory default 0% (disabled)
Controller Microprocessor
Display Digital LED, 0.1% increments
Shelves
Standard 5
Maximum 22
Construction Stainless steel, perforated
Clearance Adjustable on 1.160” centers
Dimensions (3307/3308) 17.7” x 19.9” (45.2cm x 50.5cm)
Surface area 2.4 sq. ft. per shelf
Dimensions (3310/3311) 25.7” x 19.9” (65.5cm x 50.5cm)
Surface area 3.6 sq. ft. per shelf
Loading 50 lbs. (22.7kg) stationary
Sterilization Cycle
Setpoint 140°C
Cycle time Approximately 14 hours @ 23°C, nominal line voltage
Activation Digital touch pad
Steri-Cult Incubator _____________________________________________________________________Specifications
10-1