Construction
Interior volume
3307/3308 8.2 cu. ft. (232.2 liters)
3310/3311 11.4 cu. ft. (322.8 liters)
Interior Type 304 stainless steel mirror finish, coved corners
Exterior 18 gauge cold rolled steel, powder-coated
Outer door gasket Four-sided, molded, magnetic vinyl
Inner door gasket Silicone feather
Insulation Mineral wool
Fittings
Access port 1.4” (3.6cm) with removable silicone plug
CO
2 inlet 1/4” (0.64cm) hose barbed
Drain port 3/8” (0.95cm) barbed fitting with shutoff
Electrical
115V models 115VAC, 50/60 Hz, 1PH (Operating range 90-125VAC)
Model 3307 10.5 FLA, Model 3310 11.5 FLA
Circuit breaker power switch 15A, 2 Pole (Model 3307/3310)
230V models 230VAC, 50/60 Hz, 1 PH (Operating Range 180-250VAC)
Model 3308 5.4 FLA, Model 3311 5.9 FLA
Circuit breaker power switch 10A, 2 Pole (Model 3308/3311)
Power interconnect IEC 320, male
Power inlet cord 8’ (2.4m) cord with IEC 320 female plug and country specific plug
Accessory outlet Standard, 115V/220V export, 75 watts max. (one per chamber)
Alarm contacts Deviation of temperature, CO
2, RH, power failure. N/O and N/C
Dimensions
Exterior (3307/3308) 35.0” W x 39.4” H x 27.0” F-B (88.9cm x 100.1cm x 68.6cm)
(3310/3311) 43.0” W x 39.4” H x 27.0” F-B (109.2cm x 100.1cm x 68.6cm)
Weight
Net weight Model 3307/3308 - 330 lbs. (149.7kg), Model 3310/3311 - 410 lbs. (186.0kg)
Shipping (motor) Model 3307/3308 - 400 lbs. (181.4kg), Model 3310/3311 - 480 lbs. (217.7kg)
Certifications
Refer to the Declaration of Conformity at the end of this manual
Safety Specifications
Indoor Use Only
Altitude Up to 2,000 meters
Temperature 5°C to 40°C
Humidity Maximum 80% RH for temperatures up to 31°C, decreasing linearly to 50% RH at 40°C
Mains Supply Fluctuations Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage
Installation Category II
1
Pollution Degree 2
2
Class of Equipment I
__________________________________________________________________________________________________________
1
Installation category (overvoltage category) defines the level of transient overvoltage which the instrument is designed to withstand safely. It
depends on the nature of the electricity supply and its overvoltage protection means. For example, in CAT II which is the category used for instru-
ments in installations supplied from a supply comparable to public mains such as hospital and research laboratories and most industrial laborato-
ries, the expected transient overvoltage is 2500V for a 230V supply and 1500V for a 120V supply.
2
Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2 assumes that normally
only non-conductive pollution such as dust occurs with the exception of occasional conductivity caused by condensation.
Steri-Cult Incubator _____________________________________________________________________Specifications

Table of Contents