Section 1 - Installation and Start-up
1.1 Incubator Components
Figure 1-1
Steri-Cult Incubator
Chamber Gas Sample Port - Used for sampling chamber
CO
2 content using a FYRITE or similar instrument.
Power On/Off Switch - Mains disconnect
Control Panel - Keypad, Displays & Indicators (Figure 1-2)
Fold-down Storage Area - Store manuals, markers, etc.
Leveling Feet - Used to level the unit
Note: The incubators are stackable. See Section 1.5.b.
1.2 Inventory Management System
Included in the parts bag
shipped with each unit is an invento-
ry management system. This system
consists of:
A steel plate
Five different colored magnets
Five heat-resistant shelf labels
Dry-Erase marker
Hang the plate on the inside of the outer door, using the
keyholes and hardware already installed on the door. Attach a
colored shelf label to the front edge of each shelf. The magnets
may be aligned on the plate, in order according to shelf label
color. Use the marker to record information about the product
on corresponding shelves. The magnets can be erased as needed
and rewritten.
The inventory management system magnet plate will not
withstand the heat of the sterilization cycle. Simply unhook the
whole plate from the outer door and set aside before initiating
the cycle, then reinstall afterward.
1.3 Control Panel Keys, Displays and Indicators
Figure 1-2
1. Silence - Silences the audible alarm.
2. Alarm Indicator - Light pulses on/off during an alarm
condition in the cabinet.
3. Mode Select Switch - Used to select Run, Setpoints,
Calibration and System Configuration Modes.
4. Message Center - Displays system status.
5. Mode Select Indicators -
Run: Run Menu
Settings: Set Points Menu
Calibrate: Calibration Menu
Configuration: Configuration Menu
6. Up and Down Arrows - Increments number values, tog-
gles between choices.
7. Enter - Stores the value into computer memory.
8. Heat Indicator - Lights when power is applied to the
heaters.
9. Heat Display - Displays temperature (°C)
10. Humidity Indicator - Lights when humidity is required
11. RH Display - Displays humidity (%)
12. Left and Right Arrows - Moves the operator through the
choices of the selected mode.
13. Inject Indicator - Lights when CO
2 is being injected into
the incubator chamber.
14. CO
2 Display - Displays CO2 (%)
15. Sterilization Button - Starts Sterilization Cycle
Steri-Cult Incubator ____________________________________________________________Installation and Start-Up
1 - 1