Section 5 - Alarms
5.1 Alarms
The Steri-Cult incubator alarm system is shown in the table below. When an alarm is active, the message appears in the LED
message center. Pressing Silence disables the audible alarm for the ringback period. However, the visual alarm continues until the
incubator returns to a normal condition. The alarms are momentary alarms only. When an alarm condition occurs and then returns to
normal, the incubator automatically clears the alarm condition and the message center.
Steri-Cult Incubator ___________________________________________________________________________Alarms
5 - 1
No alarm condition exists
SYSTEM OK,
CLASS 100 or
RH WAITS ON TEMP
Temperature Controller Failure
RH Conditioner Sensor Fault
CO2 > CO2 High Tracking Alarm
CO2 < CO2 Low Tracking Alarm
No water flow through the water valve
TEMP < TEMP Low Tracking Alarm
Change HEPA filter reminder-set time expired*
Tank1 is low, switch to tank 2 (Gas Guard only)
Tank2 is low, switch to tank 1 (Gas Guard only)
Tank 1 and 2 are low (Gas Guard Only)
* Visual only, no audible alarm
- All alarm delays and ringback times are ±30 seconds -
When multiple alarm conditions occur, active messages are displayed in the message center one at a time, updating at 5 second
intervals. Pressing Silence during multiple alarms causes all active alarms to be silenced and to ring back in 15 minutes.
The temperature alarms are disabled when the temperature setpoint is 10°C. The CO2 alarms are disabled when the CO2 set point
is 0.0%. RH alarms are disabled when the setpoint is 0% RH.