Incubators ______________________________________________________________________________ Preventive Maintenance
Preventive Maintenance for Steri-Cult Incubators
Refer to Manual
Section
Action Daily Weekly Monthly 3 to 6
Months
Yearly
-- Check CO
2
tank levels.
'
-- Inspect door latch, hinges and door gasket seal.
'
2 * Verify and document CO
2
, humidity and temperature
calibration, as applicable.
'
6 Thoroughly clean the interior of the incubator.
'
1.5e, f Replace HEPA and access port filters (or as needed).
'
11.3 * Replace CO2 and RH air filters.
'
5.3 Clean humidity system
'
7 Perform Sterilization Cycle as needed.
* Qualified service technicians only