Thermo iCAP 6300 - SOP: System Shut-down; Autosampler Operation Procedure

Default Icon
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
iCAP 6000 Series ICP-OES Spectrometer Hardware Manual
Page 40 of 51 Version 1.5 – March 2006
12.3.5 Shutting down the system
Place the sample tube in Deionised water and let it
pump through the system for 3 to 5 minutes.
Click on the Plasma Icon
and then click on Plasma Off.
Release the tension on the sample pump Platen.
Switch off water circulator.
12.3.6 Autosampler Operation
Click on Sequence tab at bottom of page
Select New Autosampler on the
Auto-Session Menu. This will pop
up the dialog box shown.
Select the Autosampler rack type
from the drop down box. Click on
OK
Select New this will enable you to select an already
saved method to run the Autosampler.
Input the number of samples, and a sample name
as required. Click on OK and OK again