Switch the instrument on using the power switch at the rear of the instrument (Fiyre
8:1).
Primary
~14
sohare version number and
AUTOCALB
ION
are displayed followed by
P
the time. In antmalibration procedure the i
ent selects the proper light inte
en the instrument
is
switched on, wait
1
minute for warning up.
When switching the instrument on, module
1
and its parameters (single wavelenah absorbance measurement,
no calculation, filter n:o
1,
continuous plate movement, plate analysis off and printer format) are selected.
Remove the piece of paper which has been set under the printing system during transpoffation by pressing
the
PAPER
PEED
key.
Insert the free end of the pa r roll into the slot at the rear of the instmment until a slight resistance
is
felt.
Note the correct direction of rotation ofthe roll.
l%ED
key until pa
r
is
fed through the printer. Insert the
sh&
into the pa
roll with the shaft into the guides at the rear of
the
instrument.
e
sure that there is sufllcient paper in thc printer. Never use the printer without paper.
Select the program module by pressing first the SHIFT key and then the ENTER key. Select
1
sofiware,
2
agglutination software (optional),
off software (optional) or
4
cubic spline software
Mer
entering into module,
AUTOCALIB
ION
is performed. In each module select the required
measurement mode and parameters and calculation mode and parameters. Press START key to measure. Mer
mwremenl, the results are sent to the built-in printer or via the interface.
Select the masurement mode and calculation mode before starting the measurement.
Measurement mode
is
selected with the
MEAS
MODE key. The list of
surement modes can be scro
forwards and backwards with the anow keys. The desmd mode can als
selected directly by
the
nu
keys.
ected with the
MEAS
I!
or ENTER key. The arrow and number keys are
parameters. Default v rlier selected value is selected by pressin
The following meamemenr modes can
be
selected in module
I:
Absorbances are measured at one wavelength.
2
DUAL
WAELENGTN: Absotbances arc: measured using two wavelengths:
Results of the second measurement are subtracted from results of the
first
measurement.
Absorbances of the plate are measured at two separate points in
time.
Results of the first measurement are subtracted from results of
the
second measurement.
4
TIC: Rates of absorbance change are measured
5
MULTE
WAVELENGmI-I:
Absorbanceo are measured using individual filter selection
for
each
column ofa microplate.
6
COMUTER
CONTROL: Instrument is controlled by an esternal computer.