Labsystems
Oy
IRSD
9.1
If present, adjust gains for agglutination, record filters not suitable for agglutination.
Adjust agglutination gains for each installed filter:
Starting from high lamp intensity, adjust channel gains from high to low
gain until1 a lamp intensitylchannel gain combination where all channel
levels are below MAX-SLIT-READING (9V) is found.
If the gains could not be adjusted to get channel readings between
MIN-SLIT-READING and MAX-SLIT-READING
(1V
to
9V),
the filter is not
suitable for agglutination.
10.
Select first installed filter.
11.
Switch led green unless error during initialization.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com