MULTISKAN RC and MS
5.
LINEARITY
This test can
be
used to check the linearity of the instrument.
A) Select
"5
LINEARITY and press
"ENTER-key.
B)
Enter instrument serial number,
After measurement instrument displays
user code and filter to be used.
"TEST READY
".
Press "START".
Program calculates deviation in absorbance values and percentages using linear dilution series.
After the instrument displays "TEST
READY",
values can
be
sent to an external printer by
pressing "SEND ABSn-key.
Test can be interrupted by pressing "STOP-key.
6.
PRINTER TEST
Printer test prints a test character set to the internal printer.
A) Select
"7
PRINTER and press
Test character set is printed
"ENTER-key.
Test can
be
interrupted by pressing "STOP-key.
7.
FILTER ROTATION
Filter rotation test positions each filter in ascending order starting from filter
1.
Test can be used
to adjust the filter wheel positioning.
A) Select
"7
FILTER ROTATION' and
"FILTER TEST" is displayed.
press "ENTER"-key.
B) Press any key except "STOP".
Filter number
1
is positioned.
C)
Press any key except "STOP.
Next filter is positioned.
Upper row displays the filter which is positioned and the lower row displays the next filter to be
positioned.
Test can be interrupted by pressing "STOP.
Document Release
---
Section Page
3.511352-100-~
1
.O
Labsystems
8
34