6: CONFIGURATION
Change of Configuration
By pressing "Menu" during power on or "5" key you activate the
Configuration menu.
To prevent accidental change of configuration it protected by a
code
Proceed as follows:
After pressing "5" key, the display shows: "Paßwort?"
Now type: "09122". If you entered the code correctly,
You can either load the pre-set values or change them.
Display shows: “Grundeinstellung ?”
“Ja”=0
By pressing "0" the machine loads the following values:
By pressing "Return" you skip the defaults.
Defaults values:
Language (Sprache)
1=Dutch, 2= English, 3= Italian, 4= French.
UNITS ½ mm N / in lbf (1lbf = 4.44822 N / 1 in = 25.4 mm)
LOADCELL [N] 2000
Load cell Sensitivity (“Lastzellenempfindlichkeit”) [mV/V] 3.165 (example)
Calibration factor (“Kalibrierfaktor”) [FAK] 1.000
By pressing "0" you start a Calibration program
F limit H (maximum force for Hardware) [N] 1500 [N]
repositioning speed („Rücklaufgeschwindigkeit“) [VR] 500mm/min
F (0) [N] (preload)
These settings are stored in a non volatile ROM and are re-established after
every power on.
We recommend, noting all changes on a book.