When sterilizing a large amount of liquid, it will take a longer time to raise the temperature
for articles to be sterilized. Therefore, it is recommended that the operating time should be
extended referring to the following gure:
Extended sterilization time for liquid sterilization (reference example: sterilization for 20
minutes at 121℃)
0.5L of water in a container
(for 20 minutes) + 12 minutes
1L of water in a container
(for 20 minutes) + 17 minutes
2L of water in a container
(for 20 minutes) + 28 minutes
5L of water in a container
(for 20 minutes) + 38 minutes
10L of water in a container
(for 20 minutes) + 67 minutes
<
To perform eective sterilization
>
○
The FLS-1000 controls the temperature using the temperature detected by the temperature
sensor in the chamber as an index. When more reliable sterilization is required, set the
appropriate conditions for sterilization using an external temperature sensor for articles to be
sterilized or a sterilization index body.
○