EasyManua.ls Logo

Topcon TRC-50DX - Page 50

Topcon TRC-50DX
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
48
BEFORE REQUESTING SERVICE
Operator cannot see the
patient's eye.
Illumination light is not ON. Adjust the illumination level. 33
The objective lens cap is set. Remove the objective lens cap. ---
BARRIER filter and EXCITER filter are
set to "IN".
Set BARRIER filter and EXCITER filter to
"OFF".
15
Angle changing lever stops in the mid-
dle of the "click" process.
Turn Angle changing lever to the "click"
position.
12
Diopter compensation lens selector
stops in the middle of the "click" pro-
cess.
Turn Diopter compensation lens selector
to the "click" position.
12
Internal fixation target
mount does not work.
SPLIT switch is not set to "OFF". Set SPLIT switch to "OFF". 15
Photograph of the retinal
peripheries is dark.
The distance between the patient's eye
and the instrument is longer than the
proper working distance.
Set the patient's eye closer to the instru-
ment and then take a picture.
29
Photograph of retinal cen-
ter is dark.
The pupil of the patient's eye is not fully
dilated.
Dilate the patient's eye fully. 30
Photograph is influenced
by overall flare.
The distance between the patient's eye
and instrument is shorter than the
proper working distance.
Set the patient's eye far away from the
instrument and then take a picture.
29
SMALL PUPIL switch is pressed in
angle of coverage 50°.
Press SMALL PUPIL switch to cancel
the microcoria mode.
16
Flash level is higher than necessary. Set the flash level to a low value. 33
Vague white dots are seen
through the finder and
also appear on the photo-
graph.
Tears or others adhere to the objective
lens.
Clean the objective lens. 60
Problem Condition Check Page

Table of Contents