EasyManua.ls Logo

Topcon TRC-50DX - System Classification; Purpose of Use

Topcon TRC-50DX
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
51
SPECIFICATIONS AND PERFORMANCE
SYSTEM CLASSIFICATION
Types of protection against electric shocks:
This instrument is classified as Class I equipment.
Class I equipment does not depend only on basic insulation for protection against electric
shocks, but also provides a means of connection to a protective earth system of facilities so
that metal parts that come into contact do not become conductive while the basic insulation is
in failure.
Degree of protection against electric shocks: Type B applied part
Type B applied part is the applied part complying with the specified requirements of the Stan-
dard IEC 60601-1 to provide protection against electric shock, particularly regarding allow-
able LEAKAGE CURRENT.
Degree of protection against harmful ingress of water: IPx0
The TRC-50DX has no protection against ingress of water. (The degree of protection against
harmful ingress of water defined in IEC 60529 is IPx0.)
Classification according to the method(s) of sterilization or disinfection recommended by the
manufacturer: not applicable.
The TRC-50DX has no part to be sterilized or to be disinfected.
Classification according to the degree of safety of application in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide: Equipment not suitable for use in
the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
The TRC-50DX should be used in environments where no flammable anesthetics and/or
flammable gases are present.
Classification according to the mode of operation: Continuous operation.
Continuous operation is the operation under normal load for an unlimited period, without the
specified limits of temperature being exceeded.
DIMENSIONS AND WEIGHT
Dimensions : 340mm (W) × 505mm (D) × 506~715mm (H)
Weight : 35kg (only the instrument)
PURPOSE OF USE
To observe, take pictures or record pictures of the posterior segment of the eye through the
pupil.