- 3 -
which is accessible to hospitals' PACS and HIS system and supports the available ports to
connect with the ophthalmic remote PACS system for data sharing.
1.8 Normal working conditions
[Temperature]: +10 - +40, [Relative humidity]: 40%-80%;
[Atmospheric pressure]: 760-1,060hPa.
[Power supply conditions]: Voltage: a.c.220V; Frequency: 50Hz.
1.9 Electrical safety characteristics
1) Classification by the type of electric shock protection: Class I.
2) Classification by the degree of electric shock protection: Type B, applied part.
3) Classification by the degree of liquid inlet protection stated in GB/T 4208: Conventional.
4) Classification by the degree of safety when applied together with flammable anesthetizing gas
mixed with air or with nitrous oxide: Do not use it together with the flammable anesthetizing gas
mixed with air or nitrous oxide.
5) Classification by operating mode: Continuous operation.
6) Rated voltage and frequency: 220V, 50 Hz.
7) Input power of medical electrical equipment: 200VA.
8) With/without the applied part that can shield the defibrillation discharge effects: None.
9) Whether the device has a signal output or input part: It has a signal output part.
10) Permanently/non-permanently installed equipment: Non-permanently installed equipment.
1.10 Instructions on network safety
Operation environment
Demand name
Details
Hardware conditions
CPU: 1.6GHz
RAM: 4GB
Hard disk: 120GB
Display: 10 inches
Operating system
Windows 8.1 (64 bits);
Security software:
Self-contained firewall system of Windows 8.1 (64 bits) operation system