Certain components inside the units are at high-voltage, i.e. touching these
components presents a danger to life!
Do not insert any objects into the housing
Objects inserted into the housing may result in damage to the unit or personal
injury.
Do not place any objects on top of the units
Penetrating liquids may result in a fire or electric shock.
Caution
Incorrect installation may result in extensive damage to property. Installation
should be carried out by trained personnel. In order to prevent injury to
patients and users of your systems, when installing your medical electrical
system with our products in an environment with patients, please observe the
safety requirements of EN 60601-1-1 (IEC 60601-1-1) in "Specifications for
the safety of medical electrical systems".
Take appropriate measures to particularly ensure that discharge
currents remain below the required limits: Appropriate measures:
- Disconnecting devices for signal input or output unit
- Use of a safety transformer
-Close the plug panel using the provided cover, and secure using the screws.
-Turn the switch off and then remove power cord.
Notice for users
The plug panel is closed by the cover, must not be opened by users.
Servicing information
If housing components have to be removed for servicing, this must not be
carried out in the presence of patients, the user, or other persons not involved