36
(5) Automatic mastering
The master request condition can be cleared, and the instrument can be calibrated
with the minimum and maximum masters.
Commands to instrument
Command Data type Symbol
Character string RClear master request
ASCII 52H
Character string NMinimum-mastering
ASCII 4EH
Character string XMaximum-mastering
ASCII 58H
Calibrate the instrument with the minimum and maximum masters in that order.
Data transmitted to personal computer
When automatic mastering is performed through the serial communication, one of
the data shown in the table below is returned back.
Instrument condition Data type Externally returned data
Character string @ 0 O K SP SP SP M A S SP SP SP CR LFMastering result OK
ASCII (hex) 40304F4B2020204D41532020200D0A
Character string @ 0 E R R 0 1 Z E R O M 1 CR LFZero position
ASCII (hex) 403045525230315A45524F4D310D0A
Character string @ 0 E R R 0 4 G A I N M 1 CR LFSensitivity
ASCII (hex) 403045525230344741494E4D310D0A
Character string @ 0 E R R 0 7 R E V SP M 1 CR LFMasters reversed
ASCII (hex) 40304552523037524556204D310D0A
Character string @ 0 E R R 1 0 O F F R M 1 CR LF
Mastering error
Other than above
ASCII (hex) 403045525231304F4646524D310D0A
*SP represents a space character.