User's Manual
750 50 0045 03 34
Attribute Command Description
REFx Selects the reference mode (where 'x' is 1 to 9)
1 RES2 Sets the resolution to 0.001 mm or 0.00005 in.
1 RES3 Sets the resolution to 0.01 mm or 0.0005 in.
RST
Resets the device (restarts the device) (it is managed like the "ERR"
command)
UNI? Prints the current measure unit.
1 UNI1 Enables the inch <-> milimeter conversion
1 UNI0 Disables the inch <-> milimeter conversion
VER? Prints all the firmware versions of the instrument with a label for each.
VER1? Prints the firmware version of the module without label.
VER2? Prints the firmware version of the mainboard without label.
VER3? Prints the firmware version of the sensor without label.
PCAL? Prints the last calibration date.
NCAL? Prints the next calibration date.
LCK+{seq} Locks specific keys. ('seq' = '0123456789.E' and '+' or '-')
ULK+{seq} Unlocks specific keys. ('seq' = '0123456789.E' and '+' or '-')
LN20 Disables the second line of the display.
LN21 Enables the second line of the display.
PONCE[1|0] Enables/Disables "Print once" for the probed values.
PONCE? Prints the "Print once" state.
1, 2 CHHOLD Selects the probe holder change mode.
1 CONST Selects the "constant" function.
CONST? Prints the "constant" value.
ERR? Prints information on the last error.
!ERR
Reset the error state. Used in place of a parse error, or to replace an
executed command that produced an error.
UERRE Unknown error enable
UERRD Unknown error disable + double beep