TRISAN S8 Utilities Menu F2
Miscellaneous utilities
Earth continuity 200mA. (default 1 ohm measurement threshold )
Suitable for general earth continuity testing detection of damaged earth connections.
Best for use: detecting dirty or contaminated plugs, sockets and contacts. If result greater than 1 ohm, (FAIL) clean
or replace. Test again.
PASS <1ohm.
Press START
Press START: Use setting until tester turns off
Or
Press F2: Keep setting as default
Routine Test 10A. (default 0.2 and 1 ohm measurement threshold ) Under development on TRISAN S8
Suitable for general earth continuity testing detection of damaged earth connections for new appliances
Use on END OF PRODUCTION LINE safety systems.
Best for use: Detection of broken strands in new equipment and poor crimping of terminals . If result greater than 1
ohm, (FAIL) repair or replace. Test again.
PASS <1ohm.
Press START
Press START: Use setting until tester turns off
Or
Press F2: Keep setting as default
Type Test. (default 0.2 ohm measurement threshold ) including test probe resistance
(0.1 ohm resistance allowed within appliance under test)
For use in the developmental testing of new appliances. High sensitivity to damaged earth wiring.
Best for use: Detection of poor capacity earth track wiring within the appliance to the plug. If result greater than 0.2
ohm, (FAIL) repair or replace. Test again.
PASS <0.2ohm.
Press START
Press START: Use setting until tester turn off
Or
Press F2: Keep setting as default
Measured values
For viewing failed results after every test. The need for viewing pass results is a distraction to the task. Even if the
result is close to a failed result, a competent person is not authorised to fail an appliance just below the threshold.
Threshold levels are established more than 10 fold from dangerous levels. A PASS IS A PASS alternately a FAIL IS
A FAIL for all competent personnel. Failed appliances must be removed for maintenance only at failed thresholds.
Press START
Prompt message: