JC series dual counting scale user’s guide
- 22 -
SECTION12 TECHNICAL PARAMETERS
The technical parameters accessed via the “tEch” prompt at the end of user
parameters and are password controlled to prevent unauthorised access.
These parameters set the metrology for the scales. Each scale is set independently.
The parameters will set capacity, division, decimal point position, initial zero range,
auto and manual zero range as well as factory calibration.
ACTION DISPLAYS
Enter the User Parameters section by pressing the
Pst key during the self-test when the power is
turned on. This will allow the user to set the way he
wants the scale to work by choosing specific values
from some options.
F1 off
Beep
Bk
Unit
off
Press the Smpl key until display show “tech”
tech
From the “tech” user parameter, Press the Tare key.
The scale will ask you to enter the password.
Pin
Default Password is9999. Enter “9” four times.
Press Tare.
Pin” “ ----”
Select the scale to be set up by using the
Local/Rem key.
Press the Tare to enter the technical section for that
scale.
tEch” “ LocAL” “ “
tEch” “rEmotE” “ “
Use the Smpl to select the weighing unit to be used
for setting up the parameters for the scale.
The arrow in the “Weight” window will indicate the
unit selected.
Press the Tare key to continue.
tEch” “ Unit” “ “
Press the Smpl key to scroll through the menus for
the chosen scale. User Tare to select and set
parameters and the Zero to escape as without
changing data as with the user parameter section.
Cnt
Both scales have the same
menus except that the
capacity of the local scale is
fixed and cannot be entered.
The first parameter is show internal counts.ress the
Smpl to go to next menu. Press Tare to enter.
CAp”. Scale capacity setting.
Use Smpl to select the decimal point position and
press Tare
To change the remote scale cap press CE to clear
and then enter new value followed by Tare
dESC” “ 0.00