54
SASnnn
Sets the full-scale flow rate of the analog output.
SAS Denotes set scaling factor.
nnn Sets full-scale flow rate output (range: 1 to 300 Std L/min) =
4.0 Volts output. (‘010’ denotes 10 Std L/min, leading zeroes
must be included.)
The linearized analog output can be configured for various full-scale
flow rate values. The factory default is the full scale for of the meter.
This value can be changed to improve the resolution of the analog
signal by narrowing the range of flow.
After the command is processed, an acknowledge sequence of “OK”
<CR> <LF> is sent. If the command generated an error instead of
“OK” <CR> <LF> being sent, an error code of “ERRn” <CR> <LF> will
be sent. See the end of this section for a list of possible error codes.
Use the SAVE command to permanently store the new full-scale flow
value as the new power-on default.
Example 1) SAS100 Sets scaling of analog output to 100 Std L/min
= 4.0 Volts