KANGAROO ePump™
10
Enteral Feeding Pump
Loading Pump Sets`
The KANGAROO ePump enteral feeding pump will indicate LOAD A SET in the screen’s status line if a Pump Set is not installed. A
blinking icon of a Pump Set will also appear in the upper left corner of the screen. Figure 5A shows the screen for continuous feeding
mode and Figure 5B shows the screen for intermittent feeding mode (intermittent feeding mode not available in EZMODE).
Figure 5A. LOAD A SET opening menu
screen for Continuous Feeding mode
(EZMODE not activated).
Figure 5B. LOAD A SET opening menu
screen for Intermittent Feeding mode
(EZMODE not activated).
To load a Pump Set, follow the steps below:
1. Open the blue transparent door that encloses the rotor and Pump Set loading area.
2. Load the Pump Set by holding the finger tab on the valve, and inserting the valve down into the slot on the left. (Figure 6A)
3. Grasp the MISTIC retainer end (black ring retainer) and wrap the tubing counterclockwise around the rotor. (Figure 6B)
4. Carefully pull the MISTIC retainer end upwards to position it over the slot on the right, and then lower the retainer into the slot.
(Figure 6B)
5. Press down on the valve’s finger tab to ensure proper seating.
6. Close blue transparent door.
Figure 6A.
Figure 6B.
1VNQ4FU
7BMWF
1PXFS
#VUUPO
'JOHFS5BC
.*45*$
3FUBJOFS
&OE
The display should read SET LOADED and will show one of the following opening menus, depending on the type of Pump Set
detected, and whether continuous or intermittent feeding mode has been selected.