EasyManua.ls Logo

Tyco kENDALL KANGAROO ePump - Maintenance Procedures; Side Door and Battery Door Replacement

Tyco kENDALL KANGAROO ePump
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
KANGAROO ePump
36
Enteral Feeding Pump
Top
Hinge
Pin
Top Door
Hinge
Lower Door
Hinge
Lower Hinge
Pin
Figure 35. Blue Door Replacement.
Battery Pack
See Section VIII for full instructions on replacing Battery Pack. During long-term non-usage or storage (greater than 9 months) of the
pump, battery discharge and deactivation may tend to occur. It is recommended that the battery be disconnected during this time.
The battery may need to be charged and discharged several times to restore optimal capacity.
Battery Door
To replace Battery Door, see Figure 36 and do the following:
1. Align the Battery Door Hinges with the long hinge pins.
2. Snap the hinges onto the hinge pins.
3. Screw down the door to hold in place.
Power Cord
See Section IV for initial setup, including Power Cord attachment.
Pole Clamp
See Section IV for initial setup, including attachment of the Pole Clamp to the pump.
Figure 36. Battery Door Replacement.

Table of Contents