KANGAROO ePump™
31
Enteral Feeding Pump
Figure 34. Running screen with the > 24
HRS of Pump Set Usage indicator in upper
left of the screen.
LED Array
The pump status LED array on the upper right of the pump gives a quick visual indication of the pump status, especially in darkened
rooms.
A green light status indicates normal pump operation.
A yellow light status indicates an informational or warning situation for a low battery condition, the pump in HOLDING mode, or one
of the following errors: HOLD ERROR, FEED ERROR, FLOW ERROR, and FLUSH ERROR.
A red light status indicates a critical pump error. The screen will show one of the following error messages: SYSTEM ERROR, PUMP
SET DISLODGED, or ROTOR ERROR.
Section X — Specifications and Symbols
Specifications
UL Listing
Medical Electrical Equipment
KANGAROO ePump™ Enteral Feeding Pump
(1) Classified with respect to electrical shock, fire, and mechanical hazards
in accordance with UL60601-1.
(2) Classified with respect to electrical shock, fire, mechanical and other
specified hazards in accordance with CAN/CSA C22.2 No. 601.1
Type Infusion Device
Volumetric
Pumping Mechanism
Rotary Peristaltic
Pump Sets
KANGAROO ePump MISTIC Feed-Only Set or Feed & Flush Set
Feeding Formula Delivery Rate
1-400 mL/hr in 1 mL increments