KANGAROO ePump
22
Enteral Feeding Pump
CERT INTV indicates the number of times the pump has been powered on since the last re-certification. See Section VI for details of
re-certifying the pump.
BioTech Options
Figure 22. BioTech option screen,
with Lock Setting, EZMODE, and
Communications features activated.
Lock Settings Feature — Program Locked
To “lock the settings against unauthorized changes, press4LOCK SETTINGS to get the check mark, as shown in Figure 22. The
settings can only be changed again by pressing4LOCK SETTINGS to remove the check mark. If the settings are locked and a change
is attempted, the screen will indicate PROGRAM LOCKED, and will not allow access to the parameters, Figure 16.
The pump can be PRIMED or set to RUN, HOLD, etc. when settings are locked.
EZMODE
Press4EZMODE to activate the EZMODE feeding option. A check mark will appear next to the4EZMODE button to indicate that
the feature is active. Pressing the4EZMODE button again will remove the check mark indicating that the feature is off. If the
EZMODE feature is turned on, the EZMODE icon will appear in the lower right corner of the LCD next to the battery icon (see Figure
9). See page 13 for EZMODE details.
COMMUNICATION
Press4COMMUNICATION to activate the communication feature. A check mark will appear next to the4COMMUNICATION
button to indicate that the feature is active. Pressing the4COMMUNICATION button again will remove the check mark indicating
that the feature is off. If the COMMUNICATION feature is turned on, the COMMUNICATION icon will appear in the lower right
corner of the LCD next to the battery icon (see Figure 23).
Figure 23.
Turning the COMMUNICATION feature on will initiate data transfer of various pump settings, status, errors, etc. through the serial
I/O port. If feature is desired, please contact technical support for additional information.
NOTE: If EZMODE and the COMMUNICATION feature are both activated the icons for both features will alternate.